Top

  1. Previous
    Next
    Product 2 of 2
    This product was added to our catalog on Thursday 25 February, 2016.

    Ocular MaxLight® High Mag 78D

    Price:  $0.00

    Product Description

    Design

    Each lens incorporates state of the art aspheric computer generated design and manufacturing techniques
    OI-14 through STD-LR are made of a superior optical material, which cuts fatiguing weight by 40%.
    They are guaranteed for 3 years against breakage, coating abrasion and defects in material and workmanship.
    These eight lenses also receive an exclusive hydrophobic coating, which repels water, smudges and dust, making cleaning easier.
    The 90D Standard is also available with a larger, 32mm holding ring. Product Code: OI-STD-LR
    Lenses also available in colored mounts. Contact Ocular Instruments for further information.


    Technique
    Commonly known indirect ophthalmoscopy techniques using either the slit lamp or binocular indirect ophthalmoscope should be used.
    The silver ring on the lens holder should be held toward the patient's eye during examination. It is important to recognize that this unidirectional
    design provides the best image quality possible.
    Keep the lens centered on the patient's pupil.
    Hold the lens far enough from the patient's eye so that the retinal image is the same diameter as the lens.
    Keep the illumination source as dim as possible to minimize reflections and loss of image contrast.
    Use the Ocular Lens Cleaning Cloth (OLCC) to keep lens clean and minimize glare from the lens surface.

    Cleaning
    Clean with alcohol wipe.

    Disinfecting

    Soak In:                                                                           OR
    GLUTARALDEHYDE BLEACH
    2% or 3.4% aqueous solution
    Temperature per manufacturer instructions
    Minimum exposure time = 20 minutes
    BLEACH
    10% solution mixed at:
    1 part bleach to 9 parts cool tepid water
    Recommended exposure time = 10 minutes
    Caution: To avoid damage to the lens, do not exceed recommended exposure time.
    Then:
    Rinse
    lens thoroughly to remove disinfection solution. 3 cycles of 1 minute,
    with cool or tepid water is recommended. Dry carefully and place in a
    dry storage case.


    NOTE: This lens is known to be compatible with: Asepti-Wipe, Cavi-cide, Cidex, Cidex OPA, DisCide Wipe, Enviro-cide, and Opti-Cide.
    Caution: If used on an ulcerated cornea, lens must be STERILIZED before next procedure.


    Sterilizing

    AUTOCLAVE: No
    STERRAD: No
    STERIS SYSTEM 1: No
    ETO: Yes, see ETO Parameters
    ETO Parameters: Minimum Time 1 hour, Temperature 130°F (54°C), Aeration Time 12 hours

    WARNING: Never Steam Autoclave or Boil listed lenses. Never soak in Alcohol, Acetone or Other Solvents.
    DO NOT sterilize lens in its plastic or wooden storage case.

    Specification

    Features


    Ocular Maxlight High MAG 78 indirect ophthalmoscopy diagnostic/laser aspheric lens for detailed examinations. 0.93x image mag. 1.07 laser
    spot mag. 84º static FOV, 139º dynamic FOV, 8mm working distance, 29.1mm
    clear aperture, 17g lens weight.